Users Are Unable to Save or Submit Batches with Long File Paths in Analyst® Software 1.7.2


日期: 11/01/2021
类别: Analyst Software

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Issue Description

In Analyst® software 1.7.2, users cannot save or submit an acquisition batch if the project path is longer than 126 characters. When this type of file is submitted, the following error message: “Failed to set Header data” is shown. When this file is saved, the following error message: “Failed to save BatchName” is shown.


Resolution

There is no resolution for this issue yet. The fix will be considered in future releases.