How can I avoid the following error message in Analyst® software: “Failed 5% manual integration rule. Reverting back to original peak area.”


日期: 01/31/2024
类别: Academia Omics , ProteinPilot Software , Pharma CRO

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Issue Description

The following error message appears when a manual integration is executed.
image
 


Resolution

Make sure you are in Quantitate Mode. Go to Tools, Settings, Quantitation Peak Review settings.
A screenshot of a screenshot of a window  Description automatically generated

Per default, the Manual Integration (Percent Rule) is set to “Reject manual integration if difference in new area is less than 0% of original area.” If the percentage has been changed, this can lead to an error message.