Analyst® Software Error: Failed to Write Method to Datafile Caused by Symantic Antivirus


日期: 09/18/2017
类别: Analyst Software

0 投票数
   打印    评价文章:

The Analyst® software error "Failed to write method to data file" is caused by the antivirus program, Symantec Endpoint Protection, running simultaneously during acquisition. Disabling the Sonar technology on Symantec will prevent this error. To remove Symantec's Sonar technology, the software will need to be uninstalled and reinstalled using a specific installer that omits the Sonar portion from re-installing.