Compounds Need to be Assigned to a Library to be Recognized in Cliquid® Software

A MS/MS spectra was added to a compound record in LibraryView™ software, but subsequent confirmation runs in Cliquid® software failed. The root cause of the issue was the compound in the LibraryView software database was not associated with any named library.

Cliquid® Software Does Not Show Quantitation Values for Qualitative Transitions in the Sample List

When Cliquid® software is used to submit samples, the sample list shows only the group name instead of each individual transition (e.g., only Testosterone is shown, instead of Testosterone 1 and Testosterone 2) so only one concentration from one transition can be added. To ensure that Cliquid software reads the full transition name, an underscore in between the name and number must be added (e.g., Testosterone_1 and Testosterone_2).

11/10/2021    Cliquid Software

Is SCIEX OS Software Compatible with Cliquid® Software?

No, SCIEX OS software cannot coexist with the Cliquid® software on the same PC.

11/03/2021    Cliquid Software

Is Cliquid® 3.4 Software Compatible with MPX™ 2.0 Software?

MPX™ Software is compatible with Cliquid® software and can be operated through the Cliquid software console; however, the Cliquid software Walk-up mode is not available.

04/16/2020    Cliquid Software | MPX Software

Setting Up Email Notifications for Reports in Cliquid® Software

To allow the system to deliver email notifications of reports generated in Cliquid® software, select Run Samples. In the menu option Set report options, users can choose to set up the email notifications for completed reports or indicate their preference to have a pdf version of the report emailed to them directly.

04/16/2020    Cliquid Software

Where are the Cliquid Software Report Templates Located?

The Cliquid® software report templates are saved in the folder D:\Analyst Data\Projects\Cliquid\Templates\ReportTemplates.

10/16/2019    Cliquid Software

Creating a New Report Style in Cliquid® Software

To create a new report style in Cliquid® software, go to the Administrator Tasks (Home) menu, then select Manage Report Styles and Create a Report Style.

10/10/2019    Cliquid Software

No report styles available in Cliquid 3.2

Unable to create or select report styles after a fresh install of Cliquid 3.2.

07/08/2016    Cliquid Software

Cliquid Error Retrieving the COM class factory for component with CLSID

Cliquid will run samples successfully but will be unable to process them and will report Error Retrieving the COM class factory for component with CLSID 36977493A67011Dc8Dcc00104BC5f89A at the bottom of the window.

06/02/2016    Cliquid Software

Cliquid 3.2 or MultiQuant 2.1 Installation Failure Component Visual C 2010 Runtime Librariesx86 has failed to install.

Cliquid 3.2 or MultiQuant 2.1 Installation Failure Component Visual C 2010 Runtime Librariesx86 has failed to install with the following error message Asia. Visual C 2010 Runtime Libraries x86 failed to install.