Issue Description

The error "Incorrect Argument. Hexidecimal value 0x1F is an invalid character" is received when creating reports with MultiQuant™ software.
 


Resolution

In this instance, the error was due to trailing spaces that were contained in the sample name that are not visible in the MultiQuant software results table. 

To view the spaces, open the data file (.wiff) in Analyst® software in the Explore mode and select the information icon. The trailing space in the name can be seen under File Info.

User-added image