Can the SCIEX QTRAP® or TripleQuad™ 7500 LC-MS/MS System Operate without E Lens™ Technology Active?


日期: 11/06/2020
类别: QTRAP Systems , Triple Quad Systems

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Answer

Yes, the SCIEX QTRAP® or TripleQuad™ 7500 LC-MS/MS Systems can operate without the E Lens™ Technology active. Each QTRAP and TripleQuad 7500 System comes with a blanking plug that can be inserted should there be a need to operate without the E Lens Technology. For high-flow and microflow electrospray ionization experiments, the E Lens Technology within the OptiFlow® Pro Ion Source contributes to the sensitivity improvements of the system. When using atmospheric pressure chemical ionization, the E Lens Technology cannot be used and should be replaced with the blanking plug.