LC Methods Not Opening in Analyst® Device Driver Software


日期: 06/04/2018
类别: Analyst Software

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Issue Description

Sometimes, the LC portion of a method in Analyst® Device Driver (ADD) will not open successfully; yet the associated Analyst software acquisition method file (.dam) is able to be opened, and the LC conditions can also still be viewed. This associated acquisition method can also still be run successfully in Analyst software. This occurs most prevalently (but certainly not exclusively) when copying methods from one project to another. 


Resolution

To resolve this issue, avoid copying methods involving ADD components from one Analyst project to another. (Yet, it is sometimes necessary or desirable to copy files from another project.)

Furthermore, the issue of being unable to open LC methods in ADD has also been reported when method files are not copied over from one project to another. In either case where the LC method cannot be opened in ADD, deactivating and reactivating the hardware profile will remedy the issue.