Answer

No, the concentration values for a standard or quality control samples cannot be stored in or saved to the MultiQuant™ software quantitation method.

The known concentration information must be added prior to data acquisition to the Analyst® software Build Acquisition Batch screen to the Quantitation tab or imported using the export/import functions in Analyst software (see article). This information can also be entered manually into a MultiQuant software results table and saved in the results table, but it cannot be stored in the MultiQuant software quantitation method itself.