Communication Issues when Switching Between ExionLC and Eksigent Hardware Profiles in Analyst® Software

Switching from a hardware profile that contains an ExionLC (or Shimadzu LC) system to a hardware profile with an Eksigent LC when both LCs are connected to the PC can cause Analyst® 1.6.3 software to crash, resulting in communications issues.

Unable To Change Sample Loop Volume With Analyst® Hardware Configuration For ExionLC™ AD System

After replacing the sample loop for a different volume on an ExionLC™ AD system, the volume of the loop can be changed on the LC controller panel but not within Analyst® software hardware configuration.

09/14/2022    ExionLC AD | Analyst Software

Part Number for the Tubing from the Exion/Shimadzu Autosampler to the Column Oven?

The tubing connecting the Exion/Shimadzu autosampler to the column oven is stainless steel with connectors and is coated with flexible plastic (600 mm x 0.1 mm). This part can be obtained can be obtained directly from Shimadzu (PN 228-53184-91) or from the online SCIEX store.

10/23/2017    ExionLC AC | ExionLC AD

Troubleshooting retention time shifts and variable pump pressure

Instructions for performing flow tests on Shimadzu Nexera and EXION AD HPLC pumps and troubleshooting common causes of inadequate pump flow.